Technical Information

Steinle’s Multidrop & BioTek Cassettes
Lifetime of cassettes

There are different factors to consider, which can influence the EOL:

  • Proper maintenance (cleaning, autoclaving, drying, etc.).
  • Keeping reagents, cassettes, and the environment free from particles and dust which can lead to clogs.
  • Having the cassettes under tension when not in use (better use the rest position).
  • Dispensing different reagents with one cassette (better dedicate one cassette to each liquid).
  • Dispensing reagents with properties that differ from water (ethanol, DMSO, glycerol).
  • Proper recalibration of the cassette.
  • Having a low, medium, or high throughput.
  • It is recommended to use cassettes within 1 year, securing an accurate calibration.
 

Handling Tips & Troubleshooting

For special handling tips and troubleshooting of Steinle’s Dispensing Cassettes, click here
For an individual support of your problem, please get in touch.

 

Lifetime of Steinle’s Multidrop Cassettes

Lifetime of cassettes related to total dispensing volume.
Numbers are meant as guidelines, dispensing aqueous solutions.

VersionTotal VolumeVolume Range
Standard Tube Dispensing Cassette6000 × 96 well-plates with 50µ5µl–2500µ
Best CV and accuracy at 100µ
Small Tube Dispensing Cassette150 × 1536 well-plates with 10µ0.5µl–50µ
Best CV and accuracy at 10µ

Lifetime of Steinle’s BioTek Cassettes

Lifetime of cassettes related to total dispensing volume.
Numbers are meant as guidelines, dispensing aqueous solutions.

VersionTotal VolumeVolume Range
1µ-Low Volume1,000× 384-well microplates at 5 µL per well1 µL to 50 µL
Best CV and accuracy at >2µ
5µ-Medium Volume1,000× 96-well microplates at 50 µL per well5 µL to 2,500 µL
Best CV and accuracy at >10µ
10µ-Large Volume1,000× 96-well microplates at 100 µL per well10 µL to 3,000 µL
Best CV and accuracy at >10µ

Accuracy & Precision

For Steinle’s Multidrop Dispensing Cassettes

VersionVolumeCV / Accuracy
Standard Tube Dispensing Cassette100 µLCV <1%, Accuracy <2%
Small Tube Dispensing Cassette10 µLCV <5%, Accuracy <5%

Accuracy & Precision

For Steinle’s BioTek Dispense Cassettes

VersionVolumeCV / Accuracy
1 µL – Low Volume1–50 µLCV ≤10% / ±10% @ 1 µL,
CV ≤5% / ±5% ≥ 2 µL
5 µL – Medium Volume5–2,500 µLCV ≤5% / ±4% @ 5 µL,
CV ≤2.5% / ±2% ≥ 10 µL
10 µL – Large Volume10–3,000 µLCV ≤4% / ±4% @ 10 µL,
CV ≤2% / ±2% ≥ 20 µL

Additional Information

Material
Dispensing Cassettes from Steinle Labtechnology use the same materials as conventional cassettes.
Materials that get in contact with liquids are:

Multidrop:
Polypropylene and silicone for all Standard and Small Tube plastic tipped versions .
Stainless steel, ruby and silicone for all Small Tube Metal tipped versions.
Stainless steel and silicone for the Standard Tube Metal Tip version.

BioTek:
Polypropylene and silicone for all plastic tipped versions .
Stainless steel, ruby and silicone for all Metal Tip, ruby orifice versions.
Stainless steel and silicone for all Metal Tip versions.

Warranty
Dispensing cassettes are consumables and therefore only have a limited warranty in case of malfunction.
The warranty is valid from the date you received the product, till 90 days after.

Repair or substitution is guaranteed through Steinle Labtechnology after receipt of dysfunctional product(s).
The warranty isn’t valid in case of product misuse or unusual and destructive treatment.

Calibration
All Steinle Cassettes come calibrated and ready to use including a calibration report.
A recalibration should be performed once a month or if left under tension for a prolonged period.

Autoclavability
Standard tube: 50 times (1 bar pressure at 121°C for 20 minutes).
Small tube: 10 times (1 bar pressure at 121°C for 20 minutes)

Origin of Products
Steinle Labtechnology independently develops and manufactures exclusive and compatible versions for Multidrop & BioTek Dispenser.

All products are originated quality made in Germany.
Multidrop is a registered trademark of Thermo Fisher Scientific.
BioTek is a registered trademark of BioTek Instruments, Inc.

 

 

Liquid compliance

List of dispensable liquids with all Steinle Cassettes.

If you have questions on how to deal best with a specific media, please get in touch and we will assist you to our best knowledge.

BuffersOrganic SolventsGrowth MediaBeadsProtein Solution
Hanks Hepes, cells. Cis HTRF kitAcetoneDMEM with 10% FCSPolystyrene beads0.1% BSA
IMAP detection reagentAcetonitrileKaighn’s media with 10% FBSAlpha screen beadsPlasma
IMAP 1× kinase bufferDMSO, dimethyl sulfoxideSchneider medium with 10% FBSTetra ethylene glycol carboxylated beads (0.8μm)Agar
50nM HEPES, 10% glycerol, 0.1% BSAEthanolM3 media with and without 10% FCSYttrium oxide beads
TempliPhi reagentIsopropanol
PBS bufferViscose Solutions 10% Glycerol
Tris bufferAcetic acid
0.1% Brij35 buffer
100 mM sodium phosphate w/ BSA for enzyme
20% Tris base
145 mM HEPES, 0.1% NaCl